What is a status code?
To enhance the efficiency of troubleshooting, we have incorporated status codes on the main screen of the Glunovo App.

Status code descr1ption table:
Sequence | Status code | Display |
1 | L01 | Connecting, please wait |
2 | B01 | Connection lost, please enable Bluetooth |
3 | S03 | Synchronizing, please wait |
4 | R01 | The system is being restored, please wait |
5 | R02 | The system is being restored, please wait |
6 | S01 | The system has stopped, please replace the sensor |
7 | C01 | The system power is off, please replace the sensor |
8 | T01 | The session is over, please replace the sensor |
If you see: S01, C01, and T01, you should replace the sensor and start a new session.
Please follow the steps below to obtain other status codes.
L01 System message: "Connecting, please wait."
1. Keep the screen open while the application is running and wait for 2 minutes.
2. Check the installation conditions of the transmitter to ensure it is properly installed.
Yes: Resolved
No: Fasten the transmitter
3. Upgrade to the latest version of the application
4. Restart the phone, and enable Bluetooth and the app again
5. If the above information fails to resolve the issue, please provide the phone model, App version number, phone OS version, transmitter serial number, and sensor batch number. Please send an email to: service@infinovo.com
B01: System message "Connection lost, please turn on Bluetooth"
Enable Bluetooth in the settings and reconnect to the transmitter.
S03: The system prompts "Synchronizing, please wait"
Please wait for all data to synchronize with the transmitter. Once synchronization is complete, an automatic shutdown prompt will appear. Calibration is not accepted during the data synchronization period.
R01: System message "System is recovering, please wait"
Please check if the transmitter is installed in place. If not, please snap it into place. Under no circumstances should the transmitter be removed from the sensor, as this will render the sensor inoperative.
Wait for 4 hours to confirm whether the readings have returned to normal. If not, please replace the sensor.
S01: System message "System power off. Please replace the sensor/System stopped. Please replace the sensor."
1. Check the installation of the transmitter. If it is not properly installed, please reinstall it.
2. If the transmitter is properly installed, please replace the sensor.